Accurate PV Testing: Why are NPLC and filter average designed inside SMU for PV testing?

Blog-Banner-accurate-pv-testing-smu

Table of Contents

Accurate PV Testing: Why are NPLC and filter average designed inside SMU for PV testing?

Blog-Banner-accurate-pv-testing-smu

Solar Cell I-V Characterization

A solar cell, or photovoltaic (PV), is a device that absorbs photons from a light source and then releases electrons, creating a current flow when connected to a load, thereby generating electricity. PV researchers and manufacturers are committed to achieving maximum efficiency at the lowest cost. Therefore, the analysis of electrical properties of photovoltaic cells and materials is a crucial component of the research, development, and manufacturing process. The current-voltage characteristic analysis of solar cells yields essential parameters for PV performance, including maximum current (Imax), maximum voltage (Vmax), open-circuit voltage (Voc), short-circuit current (Isc), and efficiency (Ī·).  

BL_PV_Accurate-PV-Testing-1-e1624255564148

Power measurement equipment (Source-Measure-Unit, SMU) is used in solar cell I-V characterization.

The Importance of SMU in Solar Cell I/V Testing

Many semiconductor and electronic device tests involve outputting voltage and measuring current with maximum efficiency. The overall test time comprises charging time, measuring time, discharge time, setting, and processing test time. While traditional power supplies can only output voltage or current without input capabilities, four-quadrant SMU instruments can both output and input voltages and currents, as well as measure voltage, current, or resistance. The four quadrants of SMU instruments operate in automatic input mode, enabling rapid absorption of all charges from the device under test (DUT) and wiring, effectively accelerating discharge times. Furthermore, integrating this output and measurement capability into a single instrument eliminates the need for independent digital meters (DMMs) and power supplies, improving testing time, simplifying overall testing system design, and enhancing availability. Therefore, the four-quadrant SMU (Source-Measure-Unit) is the most recommended solution for solar cell I/V testing.

Primary Noise Source in Low Current Measurement: AC Power Supply Interference

In low-current testing, noise effects on reading values must be considered. Common noise sources include:

  • Electrostatic Coupling: Interference from static electrical fields
  • Vibration/Deformation: Physical movement effects
  • Drift Current: Gradual changes in current readings
  • AC Power Interference: The most significant source of interference

BL_PV_Full-wave-rectifier-and-half-wave-rectifier-circuit.

Figure. Full-wave rectifier and half-wave rectifier circuit. Both can rectify and smooth the input AC voltage. The output is close to the ā€œDC voltageā€, but still retains the ā€œrippleā€ characteristic. This ripple creates ā€œnoiseā€ in the reading value on the meter.

Understanding AC Power Interference

How does AC power generate interfering noise? The power supply system of electrical measuring instruments primarily uses AC voltage from internal full-wave or half-wave rectifier circuits to generate “DC voltage” for related amplifying circuits, AD circuits, and other components. However, this “DC voltage” contains “ripple voltage” – a form of tiny power source instability that affects measurement circuit readings, resulting in meter reading fluctuations that constitute noise interference.

Rectifier Circuit Effects

Both full-wave and half-wave rectifier circuits can rectify and smooth input AC voltage. While the output approaches “DC voltage,” it retains “ripple” characteristics that create measurement noise. Therefore, both SMU and DMM designs must account for internal power ripple noise. Beyond refining the AC-DC conversion circuit, low-current measurements require separation or elimination of ripple noise from true signals. SMU and DMM devices incorporate built-in NPLC and average filters in their measurement functions to eliminate noise interference.

SMU Noise Elimination: NPLC/Filter/Moving-Average Functions

NPLC Filter Function

NPLC (Power Line Cycle) represents the cycle multiple of the sampling power supply, where N indicates the multiplication factor. To reduce AC power interference, measurements typically use integer multiples of the AC cycle, allowing various interferences to cancel out within one cycle.

  • NPLC Definition: The number of power line cycles representing signal sampling duration (integration time)
  • Measurement Impact: Integer multiples of NPLC produce the most accurate measurements but are limited by power line frequency (1 PLC = 60 readings/s at 60Hz, or 50 readings/s at 50Hz)

BL_PV_NPLC-filtering-and-sampling-in-the-Keithley-source-meter-manual

The above figure is a schematic diagram of NPLC filtering and sampling in the Keithley source meter manual. NPLC refers to the number of power line cycles, which represents the duration of signal sampling (or integration time). Using integer multiples of NPLC can produce the most accurate measurement, but it will be limited by the power line cycle frequency (1 PLC = 60 readings/s at 60Hz, or 50 readings/s at 50Hz).

BL_PV_the-definition-and-importance-of-NPLC-for-electrical-measurement-equipment

https://tw.tek.com/support/faqs/what-nplc-and-why-it-important

Reference: Tektronix’s website discusses NPLC’s significance in electrical measurement equipment. DC voltage, current, and resistance measurements all experience degradation due to AC power noise. Integer multiple sampling of NPLC can reduce this interference.

Integration Time and Performance Trade-offs

Using NPLC ≄1 increases data acquisition integration time and eliminates noise, improving measurement resolution and accuracy. While NPLC=100 provides maximum accuracy, it extends test time by 100x compared to NPLC=1.

BL_PV_time-axis-voltage-value

The abscissa of the figure above is the time axis, and the ordinate is the voltage value. The ā€œDCV level of the signal of interestā€ is assumed to be 115.5mV, and the blue sine wave is ā€œpower line noiseā€. When NPLC=1, the entire sampling period is just the ā€œpower line noise periodā€, so the average noise value collected is 0. This period is called the integration time.

BL_PV_time-axis-voltage-value-NPLC0.1

Let’s look at the situation when NPLC=0.1. Assuming NPLC=0.1, the time period of each measurement is integration time = 1ms. Data sampling only obtains a small part of the power line noise cycle, so the measurement result will contain noise. This is the result we see in the above figure: ā€œSignal of Interestā€ is 115.5mV, but the measured result is 116mV or 114mV.

怀怀Readers may feel very unfamiliar with the ā€œintegration timeā€ of an electric meter, and it is difficult to understand. You can understand NPLC in another way: the setting of NPLC determines the number of ā€œautomatic sampling and averagingā€ of the SMU. Assume that a reading value of SMU sampling time is fixed = 1ms. The AC power frequency is 50Hz, which means that one power line cycle is 0.02sec. When setting NPLC=1, the source meter will automatically collect (0.02sec/1ms)=20 times, and then average the 20 data to output the reading. The 20 samples are evenly distributed in a complete cycle of the power line cycle, so the average is 0, which eliminates the noise. When the single sampling time is much faster than 1ms and is close to infinitesimal, it is the concept of ā€œintegration timeā€ mentioned above.

怀怀There is a dilemma between measurement speed and performance. For example, the alternating current used is 50 Hz, and a cycle is 0.02 seconds. Therefore, if NPLC=1, the measurement sampling period is 0.02 seconds, if NPLC=10, the measurement sampling period is 0.2 seconds, and if NPLC=50, the measurement sampling period is 1 second. Only by improving the NPLC can the noise be suppressed and the AC common-mode rejection ratio can be improved. Of course, sometimes the measurement speed must be pursued, and the sampling period is less than 0.02 seconds, which will inevitably sacrifice the test performance of the electric meter. That is to say, NPLC=100 can bring the highest accuracy, but the entire test time will be 100 times that of NPLC=1.

Average Filter Function

The averaging filter operates similarly to NPLC=1, effectively eliminating power and other noise types through multiple sample averaging. Two filter types are available:

  • Moving Average: Continuously updates average using latest samples
  • Repeated Average: Averages fixed sample groups sequentially

BL_PV_built-in-Filter-function

The above figure uses Keithley’s meter as an example, which has a built-in Filter function. It provides two filter types: moving average and repeated average. It allows setting the number of filter counts and filter window. If the moving average filter is set to filter count=10, its first reading value will display the average of the 1st to 10th sampled values. The second reading shows that the sampled values from 2 to 11 will be averaged, and so on. If the filter count is set to 10 for the repeated filter, the first reading will be the average of the 1st to 10th samples, and then the reading will be output. The second reading will display the average value of the 11th to 20th samples.

NPLC Implementation

Since each source meter SMU has built-in NPLC or averaging filter functions, the user can control the NPLC or filter functions using the instrument’s front panel or a program.

BL_PV_NPLC-function-of-Keysights-SMU

The picture above shows the NPLC function of Keysight’s SMU, which explains its NPLC parameters and performance. Readers can use the buttons on the front panel to set the functions of the NPLC following the instructions in the manual.

BL_PV_mains-frequency-and-NPLC-in-the-Keithley-SMU2400-manual

The picture above is the description of how to set the mains frequency and NPLC in the Keithley SMU2400 manual. Readers who want to program control can refer to the 2400 operation manual, which has very detailed instructions. It is recommended that readers who want to program control need to have a basic understanding of the functions of the SMU and the concept of the instrument communication with the computer. This will help them get started. It is suitable for readers who have basic instrument control experience.

IVS-KA6000 SMU Control Implementation

NPLC Control through IVS-KA6000

The source meter SMU stands as the optimal I/V test tool for solar cell measurements. As the most comprehensive software for solar cell I/V measurement, IVS-KA6000 offers extensive capabilities:

  • Multiple SMU Control: Compatible with various SMU models
  • Data Analysis: Collects and analyzes current and voltage data based on set parameters
  • Parameter Analysis: Directly analyzes solar cell parameters including:
    • Short-circuit current
    • Open-circuit voltage
    • Fill factor
    • Maximum power
    • Conversion efficiency

Step-by-Step SMU Control Process

  1. Connection Verification:
    BL_PV_SMU-Connection
    • Check Source-Meter status in lower left of main screen
    • Green indicator confirms successful SMU connection
    • Enables IVS-KA6000 control capabilities
  2. Instrument Setup:
    BL_PV_SMU-Connection-Instr.-Setup
    • Access through “Instr. Setup” at top of interface
    • Enter instrument control pages
  3. Communication Settings:
    BL_PV_SMU-Connection-Config
    • Navigate to “Main” tab for SMU communication settings
    • Select SMU from drop-down menu
    • Automatic configuration of communication parameters
  4. NPLC Configuration:
    BL_PV_SMU-Connection-Source-Meter-Parameter
    • Select “Config” for advanced SMU settings
    • Access NPLC (“Number Power Line Cycles”) settings
    • Apply settings through drop-down menu
    • Confirm with “OK” button
  5. Additional Settings:
    • Configure “Current Compliance”
    • Set “2-wire/4-wire method”
    • Adjust other SMU parameters as needed
  6. Test Initiation:
    • Return to main screen
    • Begin solar cell I/V testing

Summary

The four-quadrant source meter SMU represents the optimal choice for solar cell I/V measurement. Primary noise interference typically originates from power line periodic noise, created during AC to DC power conversion. To address this, source meter SMUs incorporate built-in NPLC and filtering averaging functions for noise elimination. Users can control these functions through:

  • Panel Control: Direct SMU interface manipulation
  • Programming Control: Communication command implementation
  • Software Control: IVS-KA6000 interface utilization

Professional Expertise

  • Over a decade of experience in perovskite and organic solar cell quantum efficiency measurements since 2008
  • Comprehensive data verification and analysis software enabling rapid physical parameter extraction from quantum efficiency spectra
  • Analysis software’s physical model validated and accepted by numerous high-impact factor journals
  • Streamlined experimental validation process for journal review submissions